In order to use your TSRB 008 with Satlantic's new Satview software, your calibration file has had to be modified. The two key changes are to the coefficients of the temperature sensor, and an unused byte line. In order for the temperature sensor to work with Satview, the temperature coefficients had to be reversed. The old coefficients are in the order C3, C2, C1, C0. The new format is C0, C1, C2, and C3. The line, "SPACE NONE '' 1 AI 0 NONE," has been entered to use up a spare byte that is in the telemetry frame. These changes were only done to the final calibration file included with your shipment. The post calibration file has been left in the older format for your post processing. If you wish for all the calibration files to be left in the older format please contact Satlantic, and a new calibration file will be supplied.