» Home » HOT » Field & Laboratory Protocols


NITRATE, NITRITE and DISSOLVED ORGANIC NITROGEN

_________________________________________________________________

     SUMMARY:  Seawater is collected from known depths using 
     CTD-rosette sampling protocols.  Subsamples are carefully 
     drawn and stored in acid-washed polyethylene bottles.  
     Nitrate/nitrite is measured with an azo dye either before 
     (nitrite) or after (nitrite plus nitrate) subsamples are 
     passed through a cadmium reduction column.  Dissolved 
     organic nitrogen is determined after quantitative 
     conversion to inorganic N by exposure to UV radiation.
_________________________________________________________________


1.  Principle

    In seawater the forms of dissolved nitrogen of greatest interest 
are, in order of decreasing oxidation state:  nitrate, nitrite, ammonium 
and organic nitrogen.  All these forms of nitrogen, as well as nitrogen 
gas (N2), are biochemically interconvertible and are components of the 
biological nitrogen cycle.              

    In this method nitrate is quantitatively reduced to nitrite in a 
copperized cadmium reduction column.  The nitrite thus produced, along 
with any nitrite present in the original sample, is coupled with an 
aromatic amine, which in turn is reacted with a second aromatic amine 
to produce an azo dye.  The extinction due to the dye is then read 
spectrophotometrically. A second subsample is analyzed without prior 
reduction in order to determine the nitrite level.  Nitrate is 
calculated by difference between the [nitrate+nitrite] and nitrite 
concentrations, using standard solutions.  For surface water samples 
(<100 m) where the [nitrate+nitrite] concentration is generally <0.05 
µM, we have employed a low-level assay procedure which is based on the 
production and detection of nitrous oxides.     

    Total dissolved nitrogen (TDN) is determined by UV oxidation of the 
sample and subsequent analysis for dissolved inorganic nitrogen (DIN = 
nitrite + nitrate + ammonia).  Dissolved organic nitrogen (DON) is
computed from the relationship DON = TDN - DIN, where TDN is total 
dissolved nitrogen after UV oxidation and DIN is the sum of the 
dissolved inorganic nitrogen species before UV oxidation.  As an 
alternative to the UV oxidation method, Walsh (1989) has described 
a high-temperature (1100°C) combustion method which has been applied 
to open ocean samples collected in the North Pacific Ocean.  No 
significant differences were observed between these two procedures 
(Walsh, 1989).


2.  Precautions

    Contamination is the primary concern with these samples.  This 
is particularly true with samples collected from the euphotic zone, 
where inorganic nutrient concentrations are extremely low (<0.2 µM).  
In order to avoid contamination, sample bottles must be meticulously 
cleaned with dilute HCl and rinsed with deionized distilled water 
(DDW) before use.  Samples are stored frozen until analysis, generally 
within 1-2 weeks of sample collection.


3.  Sampling Collection and Storage

NOTE:  The currently held "dogma" in the oceanographic literature is 
that seawater samples must be processed fresh and on board ship for 
high-precision, low-level inorganic nutrient analyses (Morse et al., 
1982; Venrick and Hayward, 1985).  However, extensive results from 
automated analyses of nutrients in tropical seawaters (Ryle et al., 
1981) and the VERTEX program (D. Karl and S. Moore, unpubl. results) 
which included direct comparisons of [NO3+NO2], PO4 and SiO3 
determinations in fresh vs. frozen samples would suggest otherwise.  
Provided that caution is taken to collect and store the samples in an 
environment free of potential contamination, we found no significant 
treatment effect.  A similar conclusion was presented by Walsh et al. 
(manuscript) following the analysis of a wide range of seawater 
samples that were either analyzed fresh or frozen and stored for 
varying periods of time.  They conclude that, "Despite published and 
voiced opinions to the contrary, there appears to be no adequate 
basis either from the literature or from our experiments for 
across-the-board dismissal of high-precision nutrient analyses 
undertaken on properly stored seawater samples."  As we are not 
able to take our autoanalyzer to sea on the HOT program cruises, 
we have focussed our attention on maintaining a contamination-free 
environment during collection and storage of nutrient samples.

3.1. Sample collection

  3.1.1. Rinse the nutrient sample bottle (acid-washed, 125 ml 
         polyethylene bottle) 3 times before filling.  Fill to 
         approximately 2/3 full, tighten cap and freeze.

  3.1.2. Record cruise, cast and Niskin bottle number on the bottle 
         and data sheet.      


4.  Sample Analysis

4.1. Standard procedure

     Currently, GOFS and WOCE nutrients collected during the Hawaii 
     Ocean Time Series cruises are analyzed by the Hawaii Institute 
     of Marine Biology Analytical Facility.  Mr. Ted Walsh has 
     provided us with the following procedure for the analysis of 
     dissolved N.      

  4.1.1. Nitrate (NO3-) plus Nitrite (NO2-) 

         [NO3-+NO2-] analyses are performed on a four-channel 
         Technicon Autoanalyzer IIR continuous flow system.  The 
         automated wet chemistries generally follow the standard 
         methods of seawater analysis as given by Technicon (1977), 
         which involve:  (1) reduction of nitrate to nitrite using a 
         copperized cadmium reduction column, (2) reaction of nitrite 
         with sulfanilamide for diazotization and (3) coupling with 
         N-1-napthylethylenediamine dihydrochloride (NED) forming a 
         purple azo dye (Armstrong et al., 1967).  The dye absorbance 
         is read through a 15 mm pathlength flowcell at 550 nm.  The 
         reduction column is looped in line using a Hamilton 4-way 
         valve and can be by-passed for nitrite analysis only.  Both 
         nitrate and nitrite standards are run to check column 
         efficiency.  If speciation is desired, nitrite is determined 
         separately by omitting the reduction step.  Nitrate is 
         calculated by difference.

  4.1.2. Dissolved Organic Nitrogen (DON)

         The method for DON involves initial UV digestion of a seawater 
         sample followed by autoanalysis of the digestion products for 
         [nitrate+nitrite], as above, and ammonium using the Berthelot 
         (indophenol) method.  The modified photooxidation technique 
         (Armstrong et al., 1966) utilizes a 24 hour irradiation.  
         Details are given in Walsh (1989).  Periodic calibration 
         checks of the UV lamp efficiency are made using a dissolved 
         organic nitrogen (2,2-bypyridyl) standard.   As a general rule,
         the UV lamp is replaced after approximately 700-800 hr of use. 
         DON is calculated by difference between the sum of [nitrate+
         nitrite+ammonium] before and after UV treatment.

4.2. Low-level procedures for nitrate (NO3-) and nitrite (NO2-) by 
     chemiluminescence

     Nanomolar quantitites of nitrate and nitrite are routinely analyzed 
     using the chemiluminescence techniques of Cox (1980) and Garside 
     (1982).  This technique relies on the wet chemical reduction of 
     nitrate and nitrite to nitric oxide in a highly acidic solution 
     of sulfuric acid, ferrous ammonium sulfate, and ammonium molybdate.  
     The reduced nitric oxide is carried by an inert carrier gas (argon),
     scrubbed of acid and water vapors in a cold finger filled with 6 M 
     sodium hydroxide solution followed by an anhydrous sodium carbonate 
     filled drying tube.  The gas stream is then routed through a 
     membrane dryer and the nitric oxide is combined with ozone and 
     simultaneously exposed to a photomultiplier tube.  The nitric 
     oxide is further oxidized to excited nitrogen dioxide emitting a 
     photon as it returns to ground state, and the emitted light is 
     detected by the photomultiplier.

  4.2.1. [NO3- + NO2-] determinations 

         [NO3- + NO2-] analyses are performed on an Antek model 720 
         nitrogen oxide analyzer.  Ten ml of concentrated sulfuric 
         acid (36 N) plus 2 ml each of ferrous ammonium sulfate (4%, 
         w/v) followed by ammonium molybdate (2%, w/v) are dispensed 
         into the reaction tube.  The reaction tube is inserted into 
         the carrier gas line and the reagents degassed.  A 10 ml 
         water sample or standard solution is introduced into the 
         reaction tube with a syringe through a septum fitted to the 
         side arm of the reaction tube.  Total reaction light emission 
         is recorded using an automated integrator.

  4.2.2. NO3- determinations

         Sample analysis is conducted, as above (Chapter 7, section 
         4.2.1), except that the sample (10 ml) is pretreated with 
         0.2 ml of sulfanilamide prior to injection into the reaction 
         tube.  Since sulfanilamide quantitatively binds nitrite, the 
         integral is the result of nitrate only.  Therefore, nitrite 
         can be obtained from the difference of the two analyses.


5.  Calibration, Data Reduction and Calculations

5.1. Calibration stocks and regression standards

     The calibration of dissolved inorganic nutrients in the autoanalysis 
     of seawater samples is performed using standard solutions containing 
     N, P and Si.  A nutrient stock solution is prepared by dissolving 
     dried (65°C, 72 hours) analytical grade reagent chemicals with 
     distilled-deionized water in 1 liter glass volumetric flasks 
     containing 1 ml of chloroform.  Once dissolved, this stock solution 
     is immediately transferred into 1 liter amber polypropylene bottles 
     and stored at 4°C.  The reagent chemicals and concentrations are: 
     phosphate (KH2PO4, 1 mM), nitrate (KNO3, 5 mM) and silica (Na2SiF6, 
     4 mM).

     Working standards are prepared daily by volumetric dilutions of the 
     stock using glass pipettes and a plastic (polymethylpentene; PMP) 
     volumetric flask.  All pipettes and PMP flasks are acid-washed 
     (1 M HCl) and gravimetrically calibrated prior to use.  The daily 
     regression standards are prepared by diluting the working standard 
     with low nutrient natural seawater diluent (SWDIL).  The SWDIL is 
     filtered open ocean surface seawater that is stored in a carboy at 
     room temperature.  By using this technique all standards are 
     matrix-matched with the seawater samples and any cross-nutrient 
     interference effect should be accounted for. 

     Cross-nutrient interference and reagent contamination was evaluated 
     by preparing separate solutions, as above, but with one of the three 
     standards omitted.  Only phosphate showed a slightly measurable 
     increase (+0.014 µM) in the presence of 40 µM-NO3 and 160 µM-Si.  
     The linear regressions of the standards were applied to all seawater 
     sample peaks for calculating each batch of cruise samples.  Typical 
     correlations produced r2 values that were between 0.9999 and 0.99999.  

5.2. Blank corrections

     All seawater standard absorbance peaks were corrected for the 
     absorbance of the seawater diluent (SWDIL).  All seawater sample 
     peaks were corrected for the refractive index absorbance for each  
     unique nutrient detection system.  The refractive index corrections 
     (in apparent uM units) ranged from approximately 0.13 (for P), 0.23 
     (for N) to 2.41 (for Si), and represent the increase in absorbance 
     that is due strictly to the presence of dissolved salts in seawater 
     when compared to the distilled-deionized water baseline.  These 
     corrections are made running seawater (35 o/oo salinity) through 
     the autoanalyzer with DDW only in reagent lines.  The Levor 
     surfactant used routinely in the phosphate channel was omitted 
     from the DDW lines during the refractive index measurement because 
     Levor reacts erratically with seawater in the absence of the acidic 
     color reagent.


6.  Accuracy and Precision

    The detection limit for nitrate plus nitrite is approximately 0.03 
    uM with a coefficient of variation for field-collected replicates 
    of 0.3%.  The detection limit for DON is 0.05 with a coefficient of 
    variation of 4%.


7.  Equipment/Supplies

    Niskin bottles and rosette/CTD unit
    acid-washed, 125 ml polyethylene bottles 
    Autoanalyzer (Technicon Corp.) and accessories
    UV oxidizer unit
    nitrogen oxide analyzer (Antek model #720, operated in vacuum mode)
    reaction tube, cold finger and drying tube glassware array
          

8.  Reagents

    glass distilled deionized water (DDW)
    1 M HCl for cleaning
    concentrated H2SO4 (36 N)
    Ammonium Chloride Reagent:  Dissolve 10 g of ammonium chloride in DDW, 
     adjust pH to 8.5 with concentrated ammonium hydroxide and dilute to 
     1 liter.  Add 0.5 ml Brij-35 (Technicon No. T21-0110).
    Color Reagent:  To approximately 1500 ml of distilled water, add 200 
     ml of concentrated phosphoric acid and 20 g of sulfanilamide.  
     Dissolve completely (heat if necessary).  Add 1 g of N-1-naphthyl-
     ethylenediamine dihydrochloride and dissolve.  Dilute to 2 liters.  
     Add 1.0 ml of Brij-35. Store in a cold, dark place.  Stability: one 
     month.
    Cadmium Powder (Technicon No. T11-5063):  Use coarse cadmium powder 
     (99% pure).  Rinse the powder once or twice with a small quantity 
     of clean diethyl ether and 1 M HCl  to remove grease and dirt.  
     Follow with a DDW rinse.  Allow the metal to air-dry and store in 
     a well-stoppered bottle.
    Ferrous Ammonium Sulfate (4% w/v):  Dissolve 4 g of reagent grade 
     ferrous ammonium sulfate in 100 ml DDW.  Prepare fresh daily.
    Ammonium molybdate (2% w/v):  Dissolve 2 g reagent grade ammonium 
     molybdate in 100 ml DDW.  Prepare fresh daily.
    Sulfanilamide (1% in 10% HCl).  Dissolve 1 g reagent grade 
     sulfanilamide in 100 ml of 10% HCl.
    Sodium hydroxide (6 M):  Dissolve 240 g of reagent grade sodium 
     hydroxide and make up to 1 liter with DDW.
    Preparation of Reduction Column:  See Technicon Industrial System, 
     1977.
    Stock Standard (1000 µM):  Dissolve 0.101 g of potassium nitrate 
     in DDW and dilute to 1 liter.  Store in a dark bottle.  Add 1 ml 
     of chloroform as a preservative.
    Working Standard (50 µM):  Dilute 5 ml of stock standard in a 
     volumetric flask with DDW or seawater diluent.


9.  References

    Armstrong, F. A. J., C. R. Sterns and J. D. H. Strickland.  1967.  
    The measurement of upwelling and subsequent biological processes 
    by means of the Technicon Autoanalyzer and associated equipment. 
    Deep-Sea Research, 14, 381-389.

    Armstrong, F. A. J., P. M. Williams and J. D. H. Strickland.  
    1966.  Photo-oxidation of organic matter in seawater by ultraviolet 
    radiation, analytical and other applications. Nature, 211 481-483.

    Cox, R. D.  1980.  Determination of nitrate at the parts per 
    billion level by chemiluminescence. Analytical Chemistry, 52, 
    332-335.

    Garside, C.  1982.  A chemiluminescent technique for the 
    determination of nanomolar concentrations of nitrate and nitrite 
    in seawater. Marine Chemistry, 11, 159-167.
     
    Grasshoff, K., M. Ehrhardt, and K. Kremling.  1983.  Methods 
    of Seawater Analysis.  Verlag Chemie. 

    Morse, J. W., M. Hunt, J. Zullig, A. Mucci, and T. Mendez.  1982.  
    A comparison of techniques for preserving dissolved nutrients in 
    open ocean seawater samples. Ocean Science and Engineering, 7, 
    75-106.
                           
    Ryle, V. D., H. R. Mueller, and P. Gentien.  1981.  Automated 
    analysis of nutrients in tropical sea waters. Australian Institute 
    of Marine Science Technical Bulletin, Oceanography Series No. 3. 

    Standard Methods for the Examination of Water and Wastewater, 
    15th Edition.

    Strickland, J. D. H. and T. R. Parsons.  1972.  A Practical 
    Handbook of Seawater Analysis.  Fisheries Research Board of 
    Canada, 167.

    Technicon Industrial Systems.  1977.  Nitrate & Nitrite in 
    Water and Seawater.  Autoanalyzer II R Industrial Method No. 
    155-71W.  W. Tarrytown, New York 10591.

    Venrick, E. L. and T. L. Hayward.  1985.  Evaluation of some 
    techniques for preserving nutrients in stored seawater samples. 
    CalCOFI Report, 26, 160-168.

    Walsh, T. W.  1989.  Total dissolved nitrogen in seawater: 
    A new high temperature combustion method and a comparison to 
    photo-oxidation. Marine Chemistry, 26, 295-311.


POSTSCRIPT:  During year 1 of the HOT program, we routinely measured 
NH4+ concentrations using the standard Berthelot (indophenol) method.  
Concentrations of NH4+ were consistently at, or below, our detection 
limits (<0.05 µM) throughout the water column.  Although a new method 
has been described for low-level determinations of NH4+ in seawater 
(Brzezinski, 1988, Limnology and Oceanography, 33, 1176-1182), we have 
not yet successfully adapted this procedure for our routine determinations.  
Until that method, or a suitable alternative, is available we have decided 
to delete NH4+ measurements from our list of "standard procedures."