» Home » HOT » Field & Laboratory Protocols


LOW LEVEL NITRATE AND NITRATE BY CHEMILUMINESCENCE

_________________________________________________________________

     SUMMARY: Seawater is collected from known depths using 
     CTD-rosette sampling protocols.  Subsamples are carefully 
     drawn and stored in acid-washed polyethylene bottles.  
     Nitrate/nitrite is converted to nitric oxide by wet chemical 
     reduction in a highly acidic solution.  The nitric oxide 
     produced is measured by a chemiluminescent detector. 
_________________________________________________________________


1.  Principle

    Surface water samples (<100m) in oligotrophic waters usually 
have nitrate concentrations below the 0.03 µM detection limit of 
the Technicon Autoanalyzer.  Except for a narrow band of elevated 
nitrite around 100 m (the primary nitrite maximum), nitrite 
concentrations are also below standard detection limits.  To achieve 
high-precision high-accuracy measurements at these low levels, we 
employ the chemiluminescent method of Cox (1980) and Garside (1982).  
In this method nitrite and/or nitrate are chemically reduced to 
gaseous nitric oxide by an acidic solution of glacial acetic acid 
and sodium iodide (nitrite only) or concentrated sulfuric acid, 
ferrous ammonium sulfate and ammonium molybdate (nitrate plus 
nitrite).  The reduced nitric oxide is carried by an inert carrier 
gas (argon) through a cold finger filled with 6M sodium hydroxide, 
a drying tube filled with anhydrous sodium carbonate, and a membrane 
drier in order to remove acid and water vapors.  The gas stream is 
then routed into the chemiluminescent analyzer, where the nitric 
oxide is combined with ozone to produce a metastable nitrogen dioxide.  
The nitrogen dioxide subsequently emits a photon as it returns to 
ground state, and the emitted light is detected by a photomultiplier 
tube.  The integrated electrical signal produced by the photomultiplier 
is proportional to the content of nitrite (or nitrate plus nitrite) 
in the sample.


2.  Precautions

    As with the autoanalysis of these nutrients, contamination is the 
primary concern.  Because nanomolar concentrations are to be measured, 
all sample bottles and laboratory glassware must be meticulously cleaned 
with dilute HCl (10%) and distilled deionized water (DDW) before use, 
and both the argon carrier gas and the oxygen supplied to the ozone-
generator must be of the highest purity available.  An additional 
concern with the chemiluminescent technique is the quenching effect 
that water vapor has on the signal.  All drying and scrubbing tubes 
must be well maintained to ensure that no water enters the detector.


3.  Sample Collection and Storage

     Concerns about the potential "freezing effect" on these samples 
are heightened due to the low levels of the nutrients and the high 
precision of the measurements.  However, results from experiments 
conducted by Constantinou (unpubl.) indicate no evidence for any 
discrepancy between replicate samples analyzed immediately on ship 
and those frozen and analyzed later in the laboratory.  Samples are 
therefore collected and stored exactly as described for autoanalysis 
of these nutrients.  


4.  Analysis

    Nitrate and nitrite analyses are performed on an Antek model 720 
nitrogen oxide analyzer.  The reaction apparatus is a slight variation 
of that described by Garside (1982).  

4.1. [NO2-] determinations

     One ml sodium iodide solution (3% wt/vol) and 3 ml glacial acetic 
     acid are dispensed into the reaction tube.  The reaction tube is 
     inserted into the carrier gas line and the reagents degassed.  
     This degassing step removes any contaminating nitrite from the 
     reagents.  The extent of the degassing is monitored by the 
     detector signal.  When this step is complete, the detector is 
     reset and 10 ml of sample or standard are introduced into the 
     reaction tube with a syringe through a septum fitted to the 
     side arm of the tube.  The signal produced is integrated and 
     peak area is recorded.  The waste is then discarded, the 
     detector reset and the procedure repeated.  Generally each 
     sample or standard analysis takes about five minutes.

4.2. [NO3- + NO2-] determinations

     Ten ml concentrated sulfuric acid (36 N) followed by 2 ml 
     ferrous ammonium sulfate (4% wt/vol) and 2 ml ammonium 
     molybdate (2% wt/vol) are dispensed into the reaction tube 
     (the order of addition is important).  Unlike in the nitrite-
     only analysis, the heat generated by the dilution of the acid 
     is important for the rapid stripping of the nitric oxide 
     produced from the solution, therefore an insulating sleeve 
     is used around the reaction tube.  The reagents are degassed 
     as above, and the sample added immediately after degassing 
     is complete, in order to maintain an elevated reaction 
     temperature.  If the reaction temperature gets too low, 
     the sample may not degas rapidly enough for quantitative 
     results.  A strip chart recorder is useful for troubleshooting: 
     if asymmetrical peak tailing is observed, the reaction 
     temperature is probably not being maintained at a high enough 
     level (Garside, 1982).  The signal is integrated and recorded 
     as for the nitrite-only analysis.

4.3. [NO3-] determinations

     Our preferred method for low level nitrate determination is to 
     analyze the samples for both [NO3- + NO2-] and [NO2-] and 
     calculate [NO3-] by difference.  An alternative method is to 
     first bind the nitrite in the sample with sulfanilamide to 
     form a diazonium salt, then analyze the sample for nitrate 
     plus nitrite.  0.2 ml of a sulfanilamide solution (1% wt/vol 
     in dilute HCl) is added to the sample and allowed to react 
     with the nitrite for at least two minutes.  Then the treated 
     sample is analyzed as for [NO3- + NO2-] above.  The 
     disadvantages of using the latter method are that:  (1) a 
     small blank correction for contaminating nitrate in the 
     sulfanilamide solution must be determined and applied to the 
     results and (2) some precision is sacrificed in using the [NO3- 
     + NO2-] assay twice instead of using the [NO3- + NO2-] assay 
     once and the more precise [NO2-] assay once.  The advantage is 
     that two separate sets of analyses involving the time-consuming 
     generation of two standard curves are not required.


5.  Calibration, Data Reduction and Calculations

5.1. Calibration stocks and regression standards

     The calibration of the low level nitrate/nitrite analysis is 
     performed using standard solutions of NO3- or NO2- in DDW.  
     Dried (60°C, 72 hours) analytical grade reagent  chemicals are 
     dissolved in DDW in 500 ml acid-washed glass volumetric flasks.  
     A few drops of chloroform are added to inhibit microbial 
     activity.  The stock solutions are stored at 4°C  and discarded 
     about every three months.  The reagent chemicals and concentrations 
     are KNO3 (10 mM) and NaNO2 (10 mM).

     Working standards are prepared fresh by volumetric dilutions of 
     the stock using acid-washed glass pipettes and flasks.  Since 
     there is no salt effect in the chemiluminescent analysis, working 
     standards are all prepared using DDW.  These working standards 
     are used to generate a standard curve, and are analyzed at 
     intervals throughout the sample run in order to detect any drift 
     in the detector response.  Typical correlations produce r2 values 
     of 0.9999 or better.

     In order to maintain the accuracy of the analysis, a few certified 
     reference standards (CSK or equivalent), diluted volumetrically 
     with DDW until within the concentration range of interest, are 
     included in every sample run in case stock solutions become 
     contaminated or microbially altered during storage.

5.2. Blank corrections

     For the [NO3- + NO2-] analysis, a small blank is observed due 
     to traces of nitrate in the DDW diluent.  This blank is corrected 
     for in the calculation of standard concentrations, but is not 
     subtracted from sample results because they contain no diluent.  
     For samples, therefore, only the slope of the regression line 
     is used.  For the [NO2-] analysis, no such diluent blank is 
     observed.  If the sulfanilamide method is used for [NO3-] 
     measurements, the blank due to the sulfanilamide solution should 
     be determined and subtracted from results.


6.  Precision and Accuracy

    The detection limit for nitrate plus nitrite is approximately 
    1-2 nM.  The nitrite-only analysis produces less noisy signals; 
    standards with as little as 0.4 nM nitrite and samples with as 
    little as 0.1 nM nitrite have been detected.  Precision and 
    accuracy of the nitrate plus nitrite analysis are approximately 
    2-3 nM while for nitrite-only they are generally <2 nM.


7.  Equipment/Supplies

    Niskin bottles and rosette/CTD unit
    acid-washed 125 ml polyethylene bottles
    nitrogen oxide analyzer (Antek model #720, operated in vacuum mode)
    reaction apparatus including cold finger and drying tube
    ultra-high purity O2 and Ar, with drierite/ascarite scrubbers
    volumetric flasks and pipettes, syringe, septa, and glass beakers


8.  Reagents

    glass distilled deionized water (DDW)
    dilute HCl (10%): add 500 ml concentrated reagent grade HCl to 4500 
      ml DDW.
    glacial acetic acid
    sodium iodide (3% wt/vol): dissolve 3 g reagent grade NaI in 100 ml DDW.  
      Prepare fresh daily.
    concentrated H2SO4 (36 N)
    ferrous ammonium sulfate (4% wt/vol): dissolve 4 g reagent grade 
      ferrous ammonium sulfate in 100 ml DDW.  Prepare fresh daily.
    ammonium molybdate (2% wt/vol): dissolve 2 g reagent grade ammonium 
      molybdate in 100 ml DDW.  Prepare fresh daily.
    sulfanilamide (1% wt/vol in 10% HCl): dissolve 1 g reagent grade 
      sulfanilamide in 100 ml 10% HCl.
    sodium hydroxide (6 M): dissolve 240 g reagent grade sodium hydroxide 
      and make up to 1 l with DDW.
    nitrite stock solution (10 mM): dissolve 0.345 g dried NaNO2 in 500 ml 
      DDW.  Add 0.5 ml chloroform and store in the dark at 4°C.
    nitrate stock solution (10 mM): dissolve 0.506 g dried KNO3 in 500 ml 
      DDW.  Add 0.5 ml chloroform and store in the dark at 4°C.
    CSK (or equivalent) certified reference standards for seawater:  NO2- 
      and NO3-.


9.  References

    Cox, R. D.  1980.  Determination of nitrate at the parts per billion 
    level by chemiluminescence.  Analytical Chemistry, 52, 332-335.

    CSK Certified Seawater Reference Standards.  Sagami Chemical Research 
    Center, Sagimahara, Japan.

    Garside, C.  1982.  A chemiluminescent technique for the determination 
    of nanomolar concentrations of nitrate and nitrite in seawater.  Marine 
    Chemistry, 11, 159-167.